Crystal Instruments Spider-30R Ultra-Compact Data Acquisition System
Crystal Instruments

Spider-30RUltra-Compact Data Acquisition System

A 6-channel 24-bit ultra-compact acquisition system with 140 dB dynamic range, 102.4 kHz sampling and an aluminum enclosure shock-rated to 50 g (1000 g optional).

Manufacturer page
Phase match≤ 0.15° @ 20 kHz
Power24 VDC (±10%) / 2.5 A

Overview

The Spider-30R is an ultra-compact, high-performance data acquisition system engineered for vibration and structural measurement in challenging environments. With a rugged aluminum enclosure, sealed M-type connectors and high dynamic range, it brings laboratory-grade signal quality into a field-deployable package. Measuring just 110 x 110 x 50 mm and weighing 0.65 lb, it mounts easily on rigs, vehicles or in confined spaces. On the acquisition side, the Spider-30R offers six 24-bit sigma-delta input channels, ideal for driving two tri-axial IEPE sensors, with AC or IEPE coupling (22.5 V / 4.2 mA), a +/-10 V input range, 500 kOhm impedance, a sampling range of 0.48 Hz to 102.4 kHz and 45 kHz maximum bandwidth. Dynamic range reaches up to 140 dB, preserving accuracy from low to high amplitudes, while digital anti-aliasing provides 95 dB attenuation, amplitude match stays within +/-0.05 dB and phase match within 0.15 degrees at 20 kHz, and THD is -90 dB at 1 kHz, suiting array measurements that demand tight phase consistency. Environmental tolerance is a core strength: the enclosure is shock-tested to 50 g, an optional Extreme-Shock variant reaches 1000 g, operating temperature spans -20 to +55 degrees Celsius, and the unit is qualified against sine, random and shock vibration. Data is stored on a non-removable microSD card (4 GB default, up to 32 GB), power draw is under 6 W at 24 VDC input, and a real-time clock with +/-3.5 ppm accuracy and a 10-year backup battery supports auto power-on timers. With sealed M-type connectors and industrial ruggedization, it serves both structural vibration testing in the lab and permanently mounted predictive-maintenance front-ends.

Specifications

Crystal Instruments Spider-30R Ultra-Compact Data Acquisition System — Specifications
Phase match≤ 0.15° @ 20 kHz
Power24 VDC (±10%) / 2.5 A
Dimensions110 × 110 × 50 mm
Weight0.65 lb

Full datasheet on request.

Key features

  • 6 24-bit sigma-delta input channels, driving two tri-axial IEPE sensors
  • Up to 140 dB dynamic range for accurate low- and high-amplitude signals
  • Sampling to 102.4 kHz with 45 kHz maximum bandwidth
  • 95 dB digital anti-aliasing, amplitude match within +/-0.05 dB, phase match within 0.15 degrees
  • Aluminum enclosure shock-rated to 50 g, with an optional 1000 g Extreme-Shock variant
  • Sealed M-type connectors, -20 to +55 degrees Celsius operation, under 6 W power draw

Applications

  • Structural vibration testing and modal measurement
  • Aerospace and automotive field vibration measurement
  • Portable or embedded sensing systems
  • Long-term vibration monitoring in harsh environments
  • Permanently mounted DAQ front-end for predictive maintenance

Compare within series

ModelInput channelsSample rateADC resolutionDynamic range
Spider-30RCurrent ————
Spider-20 4———
Spider-80SG ————
Spider-80Ti ————
Spider-80Xi / 80Hi 8 (Spider-80Hi / 80Ci)——175 dBFS
Spider-HUB ————

Scroll sideways for all columns

INTERACTIVE / SIGNAL BASICS

One signal.
Different samples.

Change the sampling rate and see how sample points describe a sine wave.

Mathematical illustration: a 5 Hz sine input. Not measured equipment data or a product performance claim. Actual selection also depends on bandwidth and anti-alias filtering.

Explore signal analyzers
INPUT / 5 HzWINDOW / 1 s
Blue: original signalOrange: discrete samples

Get a quote and configuration

Tell us the test object and the standard you must meet. An engineer replies within one business day.

Related models